EBI 12-T690 data logger for measurement in EBRO steam steriliser
Po naročilu. Pošljite povpraševanje.
Delivery time: approximately 2-3 weeks
- measuring range: -20 … + 150 °C
- Memory: 4 x 25.080
- four sensors: fi 1.2 mm, radial (side) connection
Standards and directives
- (EC) 1935/2004 – materials and articles that come into contact with food
- ISO 17665 – Sterilisation of health care products
- 2014/53/EU – Radio Equipment Directive – RED
Description
The EBI 12-T690 data logger is a validation data logger designed for process validation in a steam sterilizer, recording empty chamber profiles. The Logger features flexible cable probes that allow precise positioning in the steriliser. This ensures accurate temperature monitoring during the sterilisation process.
The EBI 12-T690 features a flexible sensor that can be installed in the cavities of the sterilizer. This ensures accurate temperature monitoring even in hard-to-reach places.
The Logger has very thin temperature sensors with a diameter of 1.2 mm at the tip of the sensor, which allows very fast response times. When using the recorder in a WD, it is strongly recommended to attach the probes with silicone cable ties to ensure the stability of the sensor throughout the sterilisation process.
Who is the EBI 12-T690 suitable for?
- Healthcare facilities: the logger is ideal for monitoring sterilisation processes in hospitals, clinics, outpatient clinics and other healthcare facilities.
- Laboratories: the logger is used in a variety of laboratory environments where precise monitoring of temperature conditions during the sterilisation process is required.
- Pharmaceutical and food industry: used in the pharmaceutical and food industry where it is necessary to ensure accurate monitoring of sterilisation processes and compliance with regulatory requirements.
Technical specifications
- operating temperature: -20 … +150 °C
- measuring range : -20 … +150 °C
- temperature accuracy:
- ±0.2 °C (-20 … -0 °C)
- ±0.1 °C (0 … +120 °C),
- ±0.05 °C (+120 … +140 °C)
- ±0.1 °C (+140 … +150 °C)
- resolution: 0.01 °C
- data storage: 4 x 25,080 measurements
- Sensor: Pt1000, Class A
- Four temperature sensors
- fi 1.2 mm on flexible cables l = 600 mm
- radial connection
- sample interval: 250 ms … 24 h
- how to measure:
- infinite
- start/stop
- measurements with a fixed start
- Initial start and measurement until memory space is full
- storage temperature: -20 °C … +70 °C
- battery: lithium, 3.6 V, replaceable
- Dimensions: 48 mm x 24 mm (depending on model)
- weight: approx. 90 g
- housing: 316L stainless steel/PEEK
- Protection class: IP68
- Factory calibration certificate: 0 °C, +60 °C and +134 °C
Standards and directives
- (EC) 1935/2004 – materials and articles that come into contact with food
- ISO 17665 – Sterilisation of health care products
- 2014/53/EU – Radio Equipment Directive – RED
Software
- Set SI 2150 – Winlog.pro
- Set SI 1200 – Winlog.pro
- Set SI 3200 – Winlog.med
- Set SI 2110 – Winlog.med
- Set SI 11o0 – Winlog.med
1340-6635
Video
Read the articles:
- THE DIFFERENCE BETWEEN A HYDROMETER AND A HYGROMETER
- Measuring room temperature and humidity
- Loggers – process monitoring (cold chain, HACCP, transport, storage, process control…)
- Process monitoring, routine controls and validations
- Quality of EBRO loggers
- Data loggers for process validation
Other designations: EBI12-T690, EBI12T690, EBI-12-T690