EBI 12-T470 data logger for measurement in the EBRO steam sterilizer
Po naročilu. Pošljite povpraševanje.
Delivery time: approximately 2-3 weeks
- measuring range: -20 … + 150 °C
- Memory: 2x 50.160
- two sensors: fi 1.2 mm, axial (top) connection
- data storage: 2x 50,160 measurements
Standards and directives
- (EC) 1935/2004 – materials and articles that come into contact with food
- ISO 17665 – Sterilisation of health care products
- 2014/53/EU – Radio Equipment Directive – RED
Description
The EBI 12-T470 validation data logger is designed for thermal testing of steam steriliser processes. It is a device that includes highly flexible cable probes for precise positioning of the sensors.
The sensors are extremely thin, with a diameter of 1.2 mm at the tip of the sensor, which allows for a very fast response time.
The EBI 12-T470 is designed for use in steam sterilisers and can be used in cavities and other hard-to-reach places. The data logger is equipped with a flexible sensor that allows it to be installed in different cavities and provides accurate temperature measurements.
Who is the EBI 12-T470 suitable for?
- validation testing of steam steriliser processes in an industrial environment;
- for use in WD (wireless data logger) systems;
- use in a variety of applications requiring precise positioning and very fast response times;
- in medical institutions for testing sterilisation processes;
- in pharmaceutical companies to validate sterilisation processes;
- in sterilisation process validation laboratories;
- to monitor and control sterilisation processes on a daily basis and to ensure compliance with industry standards and regulations.
The EBI 12-T470 data logger is designed to provide accurate temperature measurements in steriliser cavities and fast, reliable data transmission. The flexible sensor can be adapted to different shapes and dimensions of sterilisers, allowing use in a variety of applications
Technical specifications
- operating temperature: -20 … +150 °C
- measuring range : -20 … +150 °C
- temperature accuracy:
- ±0.2 °C (-20 … -0 °C)
- ±0.1 °C (0 … +120 °C),
- ±0.05 °C (+120 … +140 °C)
- ±0.1°C (+140 … +150°C)
- resolution: 0.01 °C
- data storage: 2 x 50,160 measurements
- Sensor: Pt1000, Class A
- two temperature sensors fi 1.2 mm on flexible cables l = 600 or 1200 mm
- axial connection
- sample interval: 250 ms … 24 h
- how to measure:
- infinite
- start/stop
- measurements with a fixed start
- Initial start and measurement until memory space is full
- storage temperature: -20 °C … +70 °C
- battery: lithium, 3.6 V, replaceable
- dimensions: 48 mm x 24 mm
- weight: approx. 90 g
- housing: 316L stainless steel
- Protection class: IP68
- factory calibration certificate: 0 °C, +60 °C and +134 °C
Standards and directives
- (EC) 1935/2004 – materials and articles that come into contact with food
- ISO 17665 – Sterilisation of health care products
- 2014/53/EU – Radio Equipment Directive – RED
Software
- Set SI 2110 – Winlog.pro
- Set SI 1200 – Winlog.pro
- Set SI 3300 – Winlog.med
- Set SI 2100 – Winlog.med
- Set SI 2150 – Winlog.med
Video presentation:
Read the articles:
- THE DIFFERENCE BETWEEN A HYDROMETER AND A HYGROMETER
- Measuring room temperature and humidity
- Loggers – process monitoring (cold chain, HACCP, transport, storage, process control…)
- Process monitoring, routine controls and validations
- Quality of EBRO loggers
- Data loggers for process validation
Other designations: EBI12-T470, EBI12T470, EBI-12-T470