  • measuring range: -20 … + 150 °C
  • memory: 2 x 50.160
  • two sensors: fi 1.2 mm, radial (side) connection

Standards and directives

  • (EC) 1935/2004 – materials and articles that come into contact with food
  • ISO 17665 – Sterilisation of health care products
  • 2014/53/EU – Radio Equipment Directive – RED

Description

The EBI 12-T490 is a validation data logger for process validation in a steam sterilizer and batch control in production.

The Logger has quite flexible cable probes that allow precise positioning of the sensor. Very thin temperature sensors with a diameter of 1.2 mm at the tip of the sensor allow very fast response times.

When using the recorder in a WD, it is strongly recommended to attach the probes with silicone cable ties. The EBI 12-T490 is ideal for measuring temperature and humidity in sterilisation chambers, especially for accurate monitoring of the sterilisation process in cavities and other confined spaces.

 

Who is the EBI 12-T490 suitable for?

  • Healthcare: can be used in hospitals and clinics to monitor autoclave sterilisation processes, recording and documenting temperature and time data.
  • Pharmaceutical industry: can be used in pharmaceutical companies where it is necessary to monitor sterilisation processes and temperature conditions in the production of medicines.
  • Food industry: can also be used in the food industry to monitor sterilisation processes and ensure food safety.
  • Research facilities: can be used in research facilities to monitor temperature conditions in various experimental environments.

In general, the EBI 12-T490 is suitable for anyone who needs accurate monitoring of temperature and time in sterilisation processes, including autoclaves and other types of sterilisers.

 

Technical specifications

  • operating temperature: -20 °C … +150 °C
  • measuring range: -20 °C … +150 °C
  • accuracy:
    • ±0.2 °C (-20 °C … 0 °C)
    • ±0.1 °C (0 °C … +120 °C)
    • ±0.05 °C (+120 °C … +140 °C)
    • ±0.1 °C (+140 °C … +150 °C)
  • resolution: 0.01 °C
  • two sensors: fi 1.2 mm, radial (side) connection
  • data memory: 2 x 50,160 measurements
  • Sensor: Pt1000, class A
  • measuring interval: 250 ms … 24 hours
  • real-time data transmission: yes
  • storage temperature: -20 °C … +70 °C
  • measurement methods:
    • infinite measurement
    • measured at the start time
    • start immediately to the end of memory
    • start/stop of measurement
  • battery: lithium, 3.6 V, user replaceable
  • dimensions: (Ø x H) 48 mm x 24 mm
  • weight: approx. 90 g
  • Housing material: stainless steel 316 L / PEEK
  • Protection class: IP68
  • factory calibration certificate: 0 °C, +60 °C and +134 °C

 

Standards and directives

  • (EC) 1935/2004 – materials and articles that come into contact with food
  • ISO 17665 – Sterilisation of health care products
  • 2014/53/EU – Radio Equipment Directive – RED

 

Software

 

Video

Read the articles:

 

Other designations: EBI12-T490, EBI12T490, EBI-12-T490

Povpraševanje

    You may also like…