EBI 12-T671 data logger for measurement in EBRO steam sterilizer

Iskanje

Šifra: EBI 12-T671


Po naročilu. Pošljite povpraševanje.

Delivery time: approximately 2-3 weeks

  • measuring range: -20 … + 150 °C
  • Memory: 4x 25.080
  • four sensors: fi 1.2 mm, axial (top) connection

Standards and directives

  • (EC) 1935/2004 – materials and articles that come into contact with food
  • ISO 17665 – Sterilisation of health care products
  • 2014/53/EU – Radio Equipment Directive – RED

Description

The EBI 12-T671 validation data logger is designed for process validation in a steam sterilizer, recording empty chamber profiles. The Logger is equipped with highly flexible cable probes that allow precise positioning, which is particularly important when recording empty chamber profiles.

Very thin temperature sensors with a diameter of 1.2 mm at the tip of the probe allow fast response times. When using the recorder in a WD, it is recommended to attach the probes with silicone cable ties.

Using EBI 12-T671, temperature data can be accurately monitored and recorded during thermal testing of steam sterilisers, ensuring that these processes are properly validated and compliant with regulations.

 

Who is the EBI 12-T671 suitable for?

  • Manufacturers and users of steam sterilisers who need a reliable and accurate record of the temperature in the sterilisation chamber.
  • Validation institutes that test and validate sterilisation processes.
  • Pharmaceutical companies that need reliable data on sterilisation processes according to standards.
  • Hospitals and healthcare facilities that are required to carry out periodic testing and validation of sterilisation processes.
  • Laboratories that test sterilisation processes in different industries.

 

Technical specifications

  • operating temperature: -20 … +150 °C
  • measuring range : -20 … +150 °C
  • temperature accuracy:
    • ±0.2 °C (-20 … 0 °C)
    • ±0.1 °C (0 … +120 °C),
    • ±0.05 °C (+120 … +140 °C)
    • ±0.1 °C (+140 … +150 °C)
  • resolution: 0.01 °C
  • data storage: 4 x 25,080 measurements
  • Sensor: Pt1000, Class A
  • four temperature sensors fi 1.2 mm on flexible cables l = 1200 mm; axial connection
  • sample interval: 250 ms … 24 h
  • how to measure:
    • infinite
    • start/stop
    • measurements with a fixed start
    • Initial start and measurement until memory space is full
  • storage temperature: -20 °C … +70 °C
  • battery: lithium, 3.6 V, replaceable
  • Dimensions: 48 mm x 24 mm (depending on model)
  • weight: approx. 90 g
  • housing: 316L stainless steel/PEEK
  • Protection class: IP68
  • Factory calibration certificate: 0 °C, +60 °C and +134 °C

 

Standards and directives

  • (EC) 1935/2004 – materials and articles that come into contact with food
  • ISO 17665 – Sterilisation of health care products
  • 2014/53/EU – Radio Equipment Directive – RED

 

Software

Video

Read the articles:

 

Other designations: EBI12-T671, EBI12T671, EBI-12-T671