EBI 12-T220 data logger for measurement in the EBRO steam sterilizer

Iskanje

Šifra: EBI 12-T220


Po naročilu. Pošljite povpraševanje.

Delivery time: approximately 2-3 weeks

  • measuring range : -200 … +200 °C
  • resolution: 0.01 °C
  • Data storage: 100,320 measurements

Standards and Directives:

  • (EC) 1935/2004 – materials and articles that come into contact with food
  • ISO 17665 – Sterilisation of health care products – moisture
  • ISO 15883 – Washing and disinfecting machines
  • 2014/53/EU – Radio Equipment Directive RED

Description

The EBI 12-T220 data logger is an advanced data logger for process validation in steam sterilizers, disinfectant washers and endoscope washers.

It is equipped with sensors that measure the temperature, humidity and pressure inside the steriliser. The data is automatically written to the device’s internal memory, making it easy to track and analyse the data.

It is a single-channel design with a 1.5 mm flexible metal temperature probe.

Its design allows easy use and reliable real-time monitoring of sterilisation processes. This ensures the safe and efficient operation of the steam steriliser, while complying with all relevant regulations and standards.

 

Who is the EBI 12-T220 suitable for?

  • Medical and pharmaceutical industries that use steam sterilisers to sterilise disposable instruments, equipment and materials.
  • Healthcare institutions such as hospitals, health centres, laboratories, which want to ensure that all sterilisation processes are safe and effective.
  • Clinical research centres that conduct studies on sterilisation processes and need accurate data for analysis and reporting.
  • Quality engineers who are involved in quality assurance of sterilisation processes and need accurate data for analysis and reporting.
  • Technical staff to maintain and calibrate steam sterilisers.
  • Inspectors and regulatory agencies who check that sterilisation processes comply with standards and regulations.

 

Technical specifications

  • operating temperature: -90 °C … +150 °C
  • measuring range : -200 … +200 °C
  • accuracy:
    • ± 1.5 °C (-200 °C … -90 °C)
    • ± 0.5 °C (-90 °C … -40 °C)
    • ± 0.2 °C (-40 °C … 0 °C)
    • ± 0.1 °C (0 °C … +120 °C)
    • ± 0.05 °C (+120 °C … +140 °C)
    • ± 0.1 °C (+140 °C … +150 °C)
    • ± 0.5 °C (+150 °C … +200 °C)
  • resolution: 0.01 °C
  • Data storage: 100,320 measurements
  • Sensor: Pt1000, Class A
  • measuring interval: 250 ms … 24 hours
  • real-time data transmission: yes
  • measurement methods:
    • infinite measurement
    • measurement at start time
    • instant start to the end of memory
    • start/stop measurement
  • battery: lithium, 3.6 V, replaceable
  • dimensions: 48 mm x 24 mm
  • weight: approx. 90 g
  • housing: 316L stainless steel/PEEK
  • Protection class: IP68
  • factory calibration certificate: -20 °C, 0 °C, +60 °C and +134 °C

Standards and Directives:

  • (EC) 1935/2004 – materials and articles that come into contact with food
  • ISO 17665 – Sterilisation of health care products – moisture
  • ISO 15883 – Washing and disinfecting machines
  • 2014/53/EU – Radio Equipment Directive RED

 

Software

6

Video

 

We invite you to read the articles

 

Other designations: EBI12-T220, EBI12T220, EBI-12-T220

Povpraševanje