TEMPERATURE TIPALO Thermocouple (TC) wired, non-insulated, with connection connector

Iskanje

Šifra: ŽI(oznaka TC)(oznaka kabla)(dolžina)/(oznaka konektorja)


Po naročilu. Pošljite povpraševanje.

Delivery time: 3 working days or urgent

the pictures are symbolic and show type K, the colour of the insulation depends on the choice of type

  • measuring elements: thermocouples NiCr-Ni, type“K“; Fe-CuNi, type“J“; Cu-CuNi, type“T“; NiCrSi-NiSi, type“N“; PtRh-Pt“S” and“R“; NiCr-CuNi, type“E
  • colour code and accuracy class according to EN 60584
  • lengths as desired
  • Outer wire diameters: different depending on the thickness of the insulation
  • Insulation: PTFE; glass fibre, silicone, PVC, etc.
  • termination: miniature plug connector or optional
  • Sensor operating temperature T-value: varies depending on insulation
rotronicELPRO

Description

Wire thermocouples are made of insulated solid thermocouple wires and are therefore subject to the accuracy class of SIST EN 60584-1 as given in the standard, but only when used within the specified temperature range. The most commonly added connection connector is a miniature plug (flat pin). It can be implemented on LEMO or other connector formats, or on the user’s existing connectors. It is only important that the connector pins are of the same type as the thermocouple.

An uninsulated thermocouple means that the thermocouple (the joint itself) is freely visible and not protected by a protective tube. This design results in extremely fast response times. The design is simple and the elements are repairable. If it is a platinum element, the platinum is bought and sent for processing. Attention – the welding must be laser welded, without adding material, which we will be happy to take care of. This prevents errors and deviations due to contamination of the thermocouple by other substances or oxidation.

 

Wired thermocouple with connector – optional

  • measuring elements: thermocouples NiCr-Ni, type“K“; Fe-CuNi, type“J“; Cu-CuNi, type“T“; NiCrSi-NiSi, type“N“; PtRh-Pt“S” and“R“; NiCr-CuNi, type“E
  • lengths as desired
  • Outer wire diameters: different depending on the thickness of the insulation
  • Insulation: PTFE; glass fibre, silicone, PVC, etc.
  • termination: miniature plug connector or optional; for free ends, see with free ends
  • Sensor operating temperature T-value: varies depending on insulation

To achieve Class 1 accuracy, wire thermocouples may be used within the insulation range or a maximum range of -25 … +200 °C. Outside this range, the accuracy classes given are no longer valid, but use is still possible within the temperature limits of the insulation.

The wire thermocouples are laser welded in a controlled environment, so they are not contaminated with other metals or compounds.

IMPORTANT: in case of extension, use only the compensation cable, connectors and electrical terminals of the same thermocouple! All available as accessories, please ask.

Other OPTIONS at extra cost:

  • 24H: 24h production
  • portable meter or datalogger
  • ISO1: checking the temperature sensor at one selected point
  • CAL: three or more point calibration certificate, accredited
  • OV/OO: metal engraved marker plate

 

Use

Wired thermocouples are often used in combination with portable meters, calibrators, dataloggers and similar equipment. They are also popular for use in temperature control of tools, products, furnaces.

 

How to order?

From the box on the right, please send your enquiry, specifying the desired type of thermocouple, the desired length, the required temperature resistance, the quantity. We will send you a quotation with options depending on your requirements.

 

Calibration

Temperature sensors can be calibrated after manufacture and supplied with an accredited calibration certificate.

 

Custom

All temperature sensors can be made to your specifications and requirements. The most common designs are presented on the website, but in practice, production is mainly made to order. Call or email us and we will help you determine/choose the most suitable implementation for you based on your process requirements. Standard delivery time is 3 working days, urgent deliveries within 24 hours (emergency) are possible.

 

Common wired elements

NiCr-Ni, type “K”

ZIKSS3,8 silicone up to 180 °C

ZIKTT0,5X0,8 PTFE up to 250 °C

ZIKTT1,4X2,2 PTFE up to 250 °C

ZIKTT1,5X2,4 PTFE up to 250 °C

ŽIKVV0,6X1,0 glass fibre up to 400 °C

ŽIKVV1,5×2,4 glass fibre up to 400 °C

ŽIKXX1,3X2,2 special glass fibres up to 600 °C

ŽIKXX1,6X2,4 special glass fibres up to 600 °C

ŽIKXX1,6X3,2 special glass fibres up to 600 °C

ŽIKXX1,8X3,2 special glass fibres up to 600 °C

ŽIKXX2,2X3,7 special glass fibres up to 600 °C

Cu-CuNi, type “T”

ZITTT1,4X2,2 PTFE up to 250 °C

ZITTT1,6X1,0 PTFE up to 250 °C

ZITVV1,1X1,9 glass fibre up to 400 °C

 

NiCr-CuNi, type “E”

ZIETT1,4X2,2 PTFE up to 250 °C

 

Fe-CuNi, type “J”

ZIJCC1,3×1,9 capton (kapton) up to 316 °C

ZJTT1,3X2,0 PTFE up to 250 °C

ZJTT1,5X2,6 PTFE up to 250 °C

ŽIJVV0,9X1,4 glass fibre up to 400 °C

ŽIJVV1,0x1,6 glass fibre up to 400 °C

ŽIJVV1,1X1,9 glass fibre up to 400 °C

 

We invite you to read

 

Other designations: wired thermocouple with connector, wired thermocouple with connector, wired thermocouple+connector