EBI 12-TP451 data logger for EBRO benchtop autoclaves
Po naročilu. Pošljite povpraševanje.
Delivery time: okvirno 2 tedna
- measuring range: 0 °C … +150 °C, 1 … 4,000 mbar
- Temperature accuracy: ±0.1 °C, ±0.05 °C, ±0.1 °C
- resolution: 0.01 °C, 1 mbar
- data memory: 4 x 20.064 measurements
Standards and Directives:
- ISO 17665 – Sterilisation of medical products
- 2014/53/EU – Radio Equipment Directive RED
- (EC) 1935/2004 – materials and articles that come into contact with food
Description
The EBI 12-TP451 data logger for EBRO benchtop autoclaves is a validation logger for process control and validation in small sterilizers.
The Logger has 3 temperature sensors and a pressure sensor to provide accurate data on the sterilisation process. The sensors are highly flexible, allowing precise positioning in the steriliser.
The sensors have very thin tips (1.2 mm diameter), which gives them a faster response time. If the logger is used in a washer-disinfector, it is recommended to attach it with silicone cable ties.
Who is the EBI 12-TP451 suitable for?
- validation of thermal processes with steam sterilisers,
- measurement in small sterilisers,
- use in washing and disinfection processes,
- for use in medical and pharmaceutical processes, where precise positioning of sensors is crucial,
- applications where very fast sensor response times are needed.
Technical specifications
- operating temperature: 0 °C … +150 °C
- measuring range: 0 °C … +150 °C, 1 … 4,000 mbar
- temperature accuracy:
- ±0.1 °C (0 °C … +120 °C)
- ±0.05 °C (+120 °C … +140 °C)
- ±0.1 °C (+140 °C … +150 °C)
- pressure accuracy:
- ±15 mbar (1 mbar … 49 mbar)
- ±10 mbar (50 mbar … 150 mbar)
- ±15 mbar (151 mbar … 2.049 mbar)
- ±10 mbar (2.050 mbar … 2.250 mbar)
- ±15 mbar (2.241 mbar … 2.999 mbar)
- ±10 mbar (3,000 mbar … 3,250 mbar)
- ±15 mbar (3.251 mbar … 3.499 mbar)
- ±20 mbar (3,500 mbar … 4,000 mbar)
- resolution: 0.01 °C, 1 mbar
- data memory: 4 x 20.064 measurements
- sensor: pt1000, class A
- pressure sensor: piezoresistive pressure sensor
- measuring interval: 250 ms … 24 hours
- Real-time data transmission: yes
- storage temperature: -20 °C … +70 °C
- measurement methods:
- infinite measurement
- measurement at start time
- instant start to the end of memory
- start/stop measurement
- battery: lithium, 3.6 V, user replaceable
- Dimensions: (Ø x H) 48 mm x 24 mm
- weight: approx. 90 g
- Housing material: 316 L stainless steel / PEEK
- Protection class: IP68
- factory calibration certificate:
- 0 °C, +60 °C and +134 °C
- 100 mbar and 3,100 mbar (at +25 °C)
- 3,100 mbar (at +134 °C)
Standards and Directives:
- ISO 17665 – Sterilisation of medical products
- 2014/53/EU – Radio Equipment Directive RED
- (EC) 1935/2004 – materials and articles that come into contact with food
Software:
- Set SI 2150 – Winlog.pro
- Set SI 1200 – Winlog.pro
- Set SI 2110 – Winlog.med
- Set SI 1100 – Winlog.med
- Set SI 3200 – Winlog.validation
Video
You are invited to read the articles
- THE DIFFERENCE BETWEEN A HYDROMETER AND A HYGROMETER
- Room temperature measurement (external influences)
- Measuring room temperature and humidity
- How to choose between EBI 3×0 and EBI 20
- Loggers – process monitoring (cold chain, HACCP, transport, storage, process control…)
- Process monitoring, routine controls and validations
- Quality of EBRO loggers
Other designations: EBI-12-TP451, EBI 12 TP451, EBI12TP451
Povpraševanje
You may also like…
For your victories, with our solutions. Call us:
Contact us.
Call us:
+386 (0)2 62 96 720
Email us:
[email protected]