EBI 12-TP322 validation data logger for measurement in EBRO autoclaves

Iskanje

Šifra: EBI 12-TP322


Po naročilu. Pošljite povpraševanje.

Delivery time: approximately 2-3 weeks

  • measuring range temperature: 0 … +150 °C
  • measuring range pressure: 0 … 4.000 mbar
  • Memory: 3x 25,080 memory points

Standards and Directives:

  • ISO 17665 – Sterilisation of medical products
  • 2014/53/EU – Radio Equipment Directive RED
  • (EC) 1935/2004 – materials and articles that come into contact with food
  • ISO 15883 – Washing and disinfecting machines

 

Description

The EBI 12-TP322 is a validation data logger for measurement in EBRO autoclaves, used for process validation in steam sterilizers, washer disinfectors. The recorder features 2 flexible metal probes that allow precise positioning.

The ultra-thin 1.5 mm diameter temperature probes at the tip of the sensor offer extremely fast response times.

The use of a logger allows the measurement and validation of thermal processes in sterilisation. This ensures the safe and effective sterilisation of medical devices and equipment. The EBI 12-TP322 is suitable for use in a wide range of autoclaves.

 

Who is the EBI 12-TP322 suitable for?

  • Validation experts in pharmaceutical and medical companies performing thermal process tests with steam sterilisers.
  • Medical device manufacturers who want to ensure that their products are sterilised in accordance with standards and regulations.
  • Experts in laboratories that carry out thermal tests and tests on a variety of materials and products used in the pharmaceutical and medical industries.

 

Technical specifications

  • measuring range : 0 … +150 °C, 1 … 4.000 mbar
  • temperature accuracy:
    • ±0.1 °C (0 °C … +120 °C)
    • ±0.05 °C (+120 °C … +140 °C)
    • ±0.1 °C (+140 °C … +150 °C)
  • pressure accuracy:
    • ±15 mbar (1 mbar … 49 mbar)
    • ±10 mbar (50 mbar … 150 mbar)
    • ±15 mbar (151 mbar … 2.049 mbar)
    • ±10 mbar (2.050 mbar … 2.250 mbar)
    • ±15 mbar (2.241 mbar … 2.999 mbar)
    • ±10 mbar (3,000 mbar … 3,250 mbar)
    • ±15 mbar (3.251 mbar … 3.499 mbar)
    • ±20 mbar (3,500 mbar … 4,000 mbar)
  • resolution: 0.01 °C, 1 mbar
  • data storage: 3x 25.080 measurements
  • Sensor: Pt1000, Class A
  • Pressure sensor: piezoresistive
  • sample interval: 250 ms … 24 h
  • Real-time data transmission: yes
  • storage temperature: -20 °C … +70 °C
  • measurement methods:
    • infinite measurement
    • measurement at start time
    • instant start to the end of memory
    • start/stop measurement
  • battery: lithium, 3.6 V, user replaceable
  • dimensions: (Ø x H) 48 mm x 24 mm
  • weight: 90 g
  • Housing material: stainless steel 316 L / PEEK
  • Protection class: IP68
  • factory calibration certificate:
    • 0 °C, +60 °C and +134 °C
    • 100 mbar and 3,100 mbar (at +25 °C)
    • 3,100 mbar (at +134 °C)

 

Standards and Directives:

  • ISO 17665 – Sterilisation of medical products
  • 2014/53/EU – Radio Equipment Directive RED
  • (EC) 1935/2004 – materials and articles that come into contact with food
  • ISO 15883 – Washing and disinfecting machines

 

Software:

 