EBI 11-TP210 mini data logger for pasteurisation control EBRO

Iskanje

Šifra: EBI 11-TP210


Po naročilu. Pošljite povpraševanje.

Delivery time: approximately 2-3 weeks

  • the measuring area:
    • temperature: 0 °C … +140 °C
    • Pressure: 1 mbar … 8.000 mbar
  • Temperature accuracy: ±0.1 °C (0 °C … +140 °C)
  • pressure accuracy:
    • ±20 mbar (0 … 4000 mbar)
    • ±40 mbar (4.000 mbar … 8.000 mbar)
  • Resolution:
    • temperature: 0,01 °C
    • pressure: 1 mbar

Description

The EBI 11-TP210 mini data logger for pressure is the ideal solution for measuring bottle temperature and pressure. This is a miniature pressure data logger from the EBI 11 series, equipped with an M5 threaded connector for easy mounting. The compact stainless steel housing ensures its strength and longevity.

It has high accuracy up to ±20 mbar and temperature compensated pressure measurements over the entire range. The data logger has a measuring range from 1 mbar to 8,000 mbar and a temperature measuring range from 0 °C to +140 °C. Temperature accuracy is ±0.1 °C over the entire range 0 °C to +140 °C.

The device has a data storage capacity of 2 x 7,500 measurement values and a measurement interval that can be set between 1 second and 24 hours. The user can choose between different measurement modes, including infinite measurement, start time measurement, instant start to end of memory and start/stop measurement.

 

Who is the EBI 11-TP210 suitable for?

  • Measuring temperature and pressure inside bottles: small size and easy mounting on the bottle using the M5 threaded connection make this mini datalogger ideal for monitoring temperature and pressure inside bottles.
  • Pharmaceutical industry: suitable for use in the pharmaceutical industry where accurate monitoring of temperature and pressure is important for product quality and safety.
  • Laboratories: high measurement accuracy and small size make it suitable for use in laboratories where it is important that the measuring device does not interfere with the rest of the process.
  • Medical devices: used in medical devices such as infusion devices or lung ventilators where it is important to monitor temperature and pressure.
  • Food industry: used in the food industry to monitor temperature and pressure during food production and storage.

 

Technical specifications

  • operating temperature: 0 °C … +140 °C
  • the measuring area:
    • temperature: 0 °C … +140 °C
    • Pressure: 1 mbar … 8.000 mbar
  • Temperature accuracy: ±0.1 °C (0 °C … +140 °C)
  • pressure accuracy:
    • ±20 mbar (0 … 4000 mbar)
    • ±40 mbar (4.000 mbar … 8.000 mbar)
  • Resolution:
    • temperature: 0,01 °C
    • pressure: 1 mbar
  • Storage capacity: 2 x 7,500 measurement values
  • sensor: pt1000, class A
  • measuring interval: 1 s … 24 h.
  • storage temperature: -20 °C … +70 °C
  • how to measure:
    • infinite measurement
    • measurement at start time
    • instant start to the end of memory
    • start/stop measurement
  • battery: lithium, 2 x BR1225A, 3V, user replaceable
  • Dimensions: (Ø x H) 16.5 mm x 48 mm
  • weight: 45 g
  • Housing material: 316 TI stainless steel
  • Protection class: IP68
  • factory calibration certificate:
    • 0 °C, +60 °C and +134 °C
    • 100 mbar, 3,100 mbar, 6,000 at +25 °C
    • 3,100 mbar at +134 °C

 

Software (extra charge):

Choose from the following sets:

 

Standards and Directives:

  • (EC) 1935/2004 – materials and articles that come into contact with food
  • 2014/30/EU – Electromagnetic compatibility of equipment

 

Read the articles:

 

Other designations: EBI11TP210, EBI_11-TP_210, EBI-11-TP-210, EBI11_TP210