EBI 11-P111 mini data logger for pressure EBRO

Iskanje

Šifra: EBI 11-P111


Po naročilu. Pošljite povpraševanje.

  • the measuring area:
    • temperature: 0 °C … +150 °C
    • pressure: 1 mbar … 10,000 mbar
  • Temperature accuracy: ±0.1 °C (0 °C … +150 °C)
  • pressure accuracy:
    • ±15 mbar (0 … 4000 mbar)
    • ±20 mbar (4.000 mbar … 10.000 mbar)
  • Resolution:
    • temperature: 0.01 °C
    • pressure: 1 mbar

Description

The EBI 11-P111 mini data logger for pressure is an extremely accurate and reliable measuring instrument designed for pressure monitoring in a variety of applications such as benchtop sterilization, autoclaves with cassettes, washer disinfectors and endoscope washers. It allows you to accurately monitor pressure and temperature during sterilisation and washing processes.

The data logger can also be used in RDG, as well as in Sirona DAC Universal and Melag Careclave. With the Luer lock connector, it can be used universally.

It features a compact stainless steel housing that can be easily connected via a Luer lock connection. High accuracy of up to ±15 mbar allows accurate pressure measurement from 1 mbar to 4,000 mbar, with temperature-compensated pressure measurement over the entire measuring range.

It has a built-in pt1000 class A sensor for highly accurate temperature measurement from 0 °C to 150 °C. Data is recorded at intervals from 1 s to 24 hours and can be stored in the built-in memory with a capacity for 2 x 7,500 measurements.

It is equipped with long-lasting, easily replaceable batteries. The stainless steel housing provides protection against water and dust (IP68) and is designed for use in harsh environments.

 

Who is EBI 11-P111 suitable for?

  • Laboratories: ideal for laboratories carrying out sterilisation and washing procedures, it provides accurate pressure and temperature measurement in an autoclave, washer-disinfector or endoscope washer.
  • Medical facilities: its accuracy and reliability make this mini data logger ideal for use in hospitals, clinics and other medical facilities. It allows you to monitor and record pressure and temperature data during sterilisation and washing of medical instruments.
  • Pharmaceutical: useful in the pharmaceutical industry, where accuracy and reliability of measurement are crucial. It allows you to monitor pressure and temperature during sterilisation and washing processes of pharmaceuticals.
  • Manufacturing industry: suitable for use in the manufacturing industry where sterilisation and washing processes are carried out. It allows you to monitor and record pressure and temperature data during production processes.
  • Research institutions: it is also useful for research purposes where precise pressure and temperature measurement is required in various applications.
  • Technical engineers: this product is also suitable for technical engineers involved in the design and development of sterilisation and washing plants, as it allows accurate measurement of pressure and temperature in these plants.

 

Technical specifications

  • operating temperature: 0 °C … +150 °C
  • the measuring area:
    • temperature: 0 °C … +150 °C
    • pressure: 1 mbar … 10,000 mbar
  • Temperature accuracy: ±0.1 °C (0 °C … +150 °C)
  • pressure accuracy:
    • ±15 mbar (0 … 4000 mbar)
    • ±20 mbar (4.000 mbar … 10.000 mbar)
  • Resolution:
    • temperature: 0.01 °C
    • pressure: 1 mbar
  • data memory: 2 x 7,500 measurements
  • sensor: pt1000, class A
  • measuring interval: 1 s … 24 hours
  • storage temperature: -20 °C … +70 °C
  • measurement methods:
    • infinite measurement
    • measurement at start time
    • instant start to the end of memory
    • start/stop measurement
  • battery: lithium, 2 x BR1225A, 3V, user replaceable
  • Dimensions: (Ø x H) 16.5 mm x 48 mm
  • weight: 45 g
  • Housing material: 316 TI stainless steel
  • Protection class: IP68
  • factory calibration certificate:
    • 0 °C, +60 °C and +134 °C
    • 100 mbar, 3,100 mbar, 6,000 at +25 °C
    • 3,100 mbar at +134 °C

 

Software (extra charge)

Choose from the following sets:

 

Standards and Directives:

  • ISO 17665 – Sterilisation of health care products – moist heat
  • ISO 15883 – Washing and disinfecting machines
  • (EC) 1935/2004 – materials and articles that come into contact with food
  • 2014/30/EU – Electromagnetic compatibility of equipment

 

Video

 

Read the articles:

 

Other designations: EBI 11 P111, EBI11P111 , EBI-11-P111

Povpraševanje