EBI 12-TP453 validation data logger for process control and validation in small sterilizers EBRO
Po naročilu. Pošljite povpraševanje.
Delivery time: okvirno 2 tedna
- measuring range: 0 °C … +150 °C, 1 … 4,000 mbar
- resolution: 0.01 °C, 1 mbar
- data memory: 4 x 20.064 measurements
- temperature sensor: pt1000, class A
- Pressure sensor: piezoresistive
Standards and Directives:
- 2014/53/EU – Radio Equipment Directive RED
- (EC) 1935/2004 – materials and articles that come into contact with food
- ISO 17665 – Sterilisation of health care products
Description
The EBI 12-TP453 is a validation data logger used for process testing with steam sterilisers in small autoclaves. The Logger features highly flexible sensors with cables that allow precise positioning.
The Logger has 3 temperature sensors and a pressure sensor, making it ideal for measuring in smaller sterilisers. Very thin temperature sensors with a diameter of 1.2 mm at the tip of the sensor allow very fast response.
When using the logger in a washer-dryer, it is strongly recommended to fix the sensors using silicone cable ties.
Who is the EBI 12-TP453 suitable for?
- Validation of autoclave thermal processes in the laboratory workplace.
- Use in medical, pharmaceutical and biotechnology companies for measuring temperature and pressure in small sterilizers.
- Suitable for use in laboratories and research centres to check the sterility of various materials
Technical specifications
- operating temperature: 0 °C … +150 °C
- measuring range: 0 °C … +150 °C, 1 … 4,000 mbar
- temperature accuracy:
- ±0.1 °C (0 °C … +120 °C)
- ±0.05 °C (+120 °C … +140 °C)
- ±0.1 °C (+140 °C … +150 °C)
- pressure accuracy:
- ±15 mbar (1 mbar … 49 mbar)
- ±10 mbar (50 mbar … 150 mbar)
- ±15 mbar (151 mbar … 2.049 mbar)
- ±10 mbar (2.050 mbar … 2.250 mbar)
- ±15 mbar (2.241 mbar … 2.999 mbar)
- ±10 mbar (3,000 mbar … 3,250 mbar)
- ±15 mbar (3.251 mbar … 3.499 mbar)
- ±20 mbar (3,500 mbar … 4,000 mbar)
- resolution: 0.01 °C, 1 mbar
- data memory: 4 x 20.064 measurements
- Temperature sensor: Pt1000, class A
- Pressure sensor: piezoresistive
- measuring interval: 250 ms … 24 hours
- Real-time data transmission: yes
- storage temperature: -20 °C … +70 °C
- measurement methods:
- infinite measurement
- measurement at start time
- instant start to the end of memory
- start/stop measurement
- battery: lithium, 3.6 V, user replaceable
- Dimensions: (Ø x H) 48 mm x 24 mm
- weight: 90 g
- Housing material: 316 L stainless steel / PEEK
- Protection class: IP68
- factory calibration certificate:
- 0 °C, +60 °C and +134 °C
- 100 mbar and 3,100 mbar (at +25 °C)
- 3,100 mbar (at +134 °C)
Standards and Directives:
- 2014/53/EU – Radio Equipment Directive RED
- (EC) 1935/2004 – materials and articles that come into contact with food
- ISO 17665 – Sterilisation of health care products
Software:
- Set SI 2150 – Winlog.pro
- Set SI 1200 – Winlog.pro
- Set SI 2110 – Winlog.med
- Set SI 1100 – Winlog.med
- Set SI 3200 – Winlog.validation
Video
Read the articles:
- THE DIFFERENCE BETWEEN A HYDROMETER AND A HYGROMETER
- Room temperature measurement (external influences)
- Measuring room temperature and humidity
- How to choose between EBI 3×0 and EBI 20
- Loggers – process monitoring (cold chain, HACCP, transport, storage, process control…)
- Process monitoring, routine controls and validations
- Quality of EBRO loggers
Other designations: EBI-12-TP453, EBI 12 TP453, EBI12TP453, EBI_12_TP453
Povpraševanje
You may also like…
For your victories, with our solutions. Call us:
Contact us.
Call us:
+386 (0)2 62 96 720
Email us:
[email protected]