EBI 12-TP231-EX universal validation data logger EBRO

Iskanje

Šifra: EBI 12-TP231-EX


Po naročilu. Pošljite povpraševanje.

Delivery time: approximately 2-3 weeks

  • operating temperature: 0 °C … +150 °C
  • ATEX approval: yes
  • permissible environmental conditions in the danger zone:
    • 0 °C … +85 °C
    • 100 Pa … 400 kPa
  • the measuring area:
    • 0 °C … +150 °C
    • 1 … 4,000 mbar

Description

The EBI 12-TP231-EX is a universal validation data logger that can also be used in hazardous areas (EX). The Logger is designed for process control and validation in steam sterilizers, washer-disinfectors. It contains an external, rigid, axial temperature sensor with a diameter of 1.95 mm.

It is characterised by its high precision and ease of use. In addition to validation, it can also be used for routine control in medical premises. Using this recorder, data can be collected that is important for monitoring and improving sterilisation processes and ensuring the safety and reliability of medical devices.

 

For whom the EBI 12-TP231-EX is suitable

  • Companies working with hazardous substances and operating in environments where there is a risk of explosion
  • Technicians and engineers working in the field of validation of autoclave and WD sterilisation processes in EX-zones
  • Leaders in companies that ensure the quality, safety and efficiency of sterilisation processes in medical facilities where autoclaves and WDs are used.

 

Technical specifications

  • operating temperature: 0 °C … +150 °C
  • ATEX approval: yes
  • permissible environmental conditions in the danger zone:
    • 0 °C … +85 °C
    • 100 Pa … 400 kPa
  • the measuring area:
    • 0 °C … +150 °C
    • 1 … 4,000 mbar
  • temperature accuracy:
    • ±0.1 °C (0 °C … +120 °C)
    • ±0.05 °C (+120 °C … +140 °C)
    • ±0.1 °C (+140 °C … +150 °C)
  • pressure accuracy:
    • ±15 mbar (1 mbar … 49 mbar)
    • ±10 mbar (50 mbar … 150 mbar)
    • ±15 mbar (151 mbar … 2.049 mbar)
    • ±10 mbar (2.050 mbar … 2.250 mbar)
    • ±15 mbar (2.241 mbar … 2.999 mbar)
    • ±10 mbar (3,000 mbar … 3,250 mbar)
    • ±15 mbar (3.251 mbar … 3.499 mbar)
    • ±20 mbar (3,500 mbar … 4,000 mbar)
  • resolution: 0.01 °C, 1 mbar
  • data memory: 2x 33,440 measurements
  • Temperature sensor: Pt1000, class A
  • pressure sensor: piezoresistive pressure sensor
  • measuring interval: 250 ms … 24 h
  • Real-time data transmission: yes
  • storage temperature: -20 °C … +70 °C
  • measurement methods:
    • infinite measurement
    • measurement at start time
    • instant start to the end of memory
    • start/stop measurement
  • battery: lithium, 3.6 V, user replaceable
  • dimensions: (Ø x H) 48 mm x 24 mm
  • weight: 90 g
  • Housing material: stainless steel 316 L / PEEK
  • Protection class: IP68
  • factory calibration certificate:
    • 0 °C, +60 °C and +134 °C
    • 100 mbar and 3,100 mbar (for +25 °C)
    • 3,100 mbar (at +134 °C)

 

Standards and Directives:

  • ISO 17665 – Sterilisation of medical products
  • (EC) 1935/2004 – materials and articles that come into contact with food
  • 2014/34/EUATEX Directive
  • ISO 15883 – Disinfectors For Washing Machines

ATEX type

  • II1G Ex ia IIC T4 Ga

 

Software

Video

 

Read the articles:

 

Other designations: EBI12-TH231, EBI12-TH231, EBI12TH231, EBI12TH231, EBI-12-TH231, EBI-12-TH231, EBI-12-TH231, EBI12-TH231-EX, EBI12-TH231EX, EBI12TH231EX, EBI-12TH231EX, EBI-12-TH231-EX

Povpraševanje