EBI 12-TP226 validation data logger for EBRO steam sterilisers
Po naročilu. Pošljite povpraševanje.
Delivery time: approximately 2-3 weeks
- measuring range temperature: 0 … +150 °C
- measuring range pressure: 0 … 4.000 mbar
- Memory: 2 x 33,440 memory points
Standards and Directives:
- ISO 17665 – Sterilisation of medical products
- 2014/53/EU – Radio Equipment Directive RED
- (EC) 1935/2004 – materials and articles that come into contact with food
Description
The EBI 12-TP226 is a validation data logger for temperature and pressure measurement in steam sterilisers. The logger is characterised by flexible metal probes that provide accurate temperature measurements even in hard-to-reach places.
The probe can also be used to measure pressure. The very thin, flexible temperature sensors with a diameter of 1.5 mm allow extremely fast response times.
A data logger is often used to measure the temperature in the steriliser drain line while simultaneously measuring the pressure.
Who is the EBI 12-TP231 suitable for?
- Companies and organisations that perform steam sterilisation and want to validate and control the sterilisation process.
- Professionals and technicians who carry out validation and control of the steam sterilisation process in medical, pharmaceutical and laboratory applications.
Technical specifications
- operating temperature: 0 °C … +150 °C
- measuring range: 0 °C … +150 °C, 1 … 4,000 mbar
- temperature accuracy:
- ±0.1 °C (0 °C … +120 °C)
- ±0.05 °C (+120 °C … +140 °C)
- ±0.1 °C (+140 °C … +150 °C)
- pressure accuracy:
- ±15 mbar (1 mbar … 49 mbar)
- ±10 mbar (50 mbar … 150 mbar)
- ±15 mbar (151 mbar … 2.049 mbar)
- ±10 mbar (2.050 mbar … 2.250 mbar)
- ±15 mbar (2.241 mbar … 2.999 mbar)
- ±10 mbar (3,000 mbar … 3,250 mbar)
- ±15 mbar (3.251 mbar … 3.499 mbar)
- ±20 mbar (3,500 mbar … 4,000 mbar)
- resolution: 0.01 °C, 1 mbar
- data memory: 2 x 33,440 measurements
- sensor: pt1000, class A
- pressure sensor: piezoresistive pressure sensor
- measuring interval: 250 ms … 24 hours
- Real-time data transmission: yes
- storage temperature: -20 °C … +70 °C
- measurement methods:
- infinite measurement
- measurement at start time
- instant start to the end of memory
- start/stop measurement
- battery: lithium, 3.6 V, user replaceable
- Dimensions: (Ø x H) 48 mm x 24 mm
- weight: approx. 90 g
- Housing material: 316 L stainless steel / PEEK
- Protection class: IP68
- factory calibration certificate:
- 0 °C, +60 °C and +134 °C
- 100 mbar and 3,100 mbar (at +25 °C)
- 3,100 mbar (at +134 °C)
Standards and Directives:
- ISO 17665 – Sterilisation of medical products
- 2014/53/EU – Radio Equipment Directive RED
- (EC) 1935/2004 – materials and articles that come into contact with food
Software:
- Set SI 2150 – Winlog.pro
- Set SI 1200 – Winlog.pro
- Set SI 2110 – Winlog.med
- Set SI 2100 – Winlog.med
- Set SI 3200 – Winlog.validation
Video
Read the articles:
- THE DIFFERENCE BETWEEN A HYDROMETER AND A HYGROMETER
- Measuring room temperature and humidity
- Loggers – process monitoring (cold chain, HACCP, transport, storage, process control…)
- Process monitoring, routine controls and validations
- Quality of EBRO loggers
- Data loggers for process validation
Other designations: EBI12-TP226, EBI12-TP226, EBI12TP226, EBI12TP226, EBI-12-TP226, EBI-12-TP226, EBI-12-TP226