  • measuring range temperature: 0 … +150 °C
  • measuring range pressure: 0 … 4.000 mbar
  • Memory: 2 x 33,440 memory points

Standards and Directives:

  • ISO 17665 – Sterilisation of medical products
  • 2014/53/EU – Radio Equipment Directive RED
  • (EC) 1935/2004 – materials and articles that come into contact with food

Description

The EBI 12-TP226 is a validation data logger for temperature and pressure measurement in steam sterilisers. The logger is characterised by flexible metal probes that provide accurate temperature measurements even in hard-to-reach places.

The probe can also be used to measure pressure. The very thin, flexible temperature sensors with a diameter of 1.5 mm allow extremely fast response times.

A data logger is often used to measure the temperature in the steriliser drain line while simultaneously measuring the pressure.

 

Who is the EBI 12-TP231 suitable for?

  • Companies and organisations that perform steam sterilisation and want to validate and control the sterilisation process.
  • Professionals and technicians who carry out validation and control of the steam sterilisation process in medical, pharmaceutical and laboratory applications.

 

Technical specifications

  • operating temperature: 0 °C … +150 °C
  • measuring range: 0 °C … +150 °C, 1 … 4,000 mbar
  • temperature accuracy:
    • ±0.1 °C (0 °C … +120 °C)
    • ±0.05 °C (+120 °C … +140 °C)
    • ±0.1 °C (+140 °C … +150 °C)
  • pressure accuracy:
    • ±15 mbar (1 mbar … 49 mbar)
    • ±10 mbar (50 mbar … 150 mbar)
    • ±15 mbar (151 mbar … 2.049 mbar)
    • ±10 mbar (2.050 mbar … 2.250 mbar)
    • ±15 mbar (2.241 mbar … 2.999 mbar)
    • ±10 mbar (3,000 mbar … 3,250 mbar)
    • ±15 mbar (3.251 mbar … 3.499 mbar)
    • ±20 mbar (3,500 mbar … 4,000 mbar)
  • resolution: 0.01 °C, 1 mbar
  • data memory: 2 x 33,440 measurements
  • sensor: pt1000, class A
  • pressure sensor: piezoresistive pressure sensor
  • measuring interval: 250 ms … 24 hours
  • Real-time data transmission: yes
  • storage temperature: -20 °C … +70 °C
  • measurement methods:
    • infinite measurement
    • measurement at start time
    • instant start to the end of memory
    • start/stop measurement
  • battery: lithium, 3.6 V, user replaceable
  • Dimensions: (Ø x H) 48 mm x 24 mm
  • weight: approx. 90 g
  • Housing material: 316 L stainless steel / PEEK
  • Protection class: IP68
  • factory calibration certificate:
    • 0 °C, +60 °C and +134 °C
    • 100 mbar and 3,100 mbar (at +25 °C)
    • 3,100 mbar (at +134 °C)

 

Standards and Directives:

  • ISO 17665 – Sterilisation of medical products
  • 2014/53/EU – Radio Equipment Directive RED
  • (EC) 1935/2004 – materials and articles that come into contact with food

 

Software: