  • measuring range: 0 … +125 °C, 1 µS/cm … 2.000 µS/cm
  • resolution: 0.01 °C, 0.1 µS/cm
  • data storage: 2x 50,160 measurements
  • Sensor: Pt1000, conduction electrode

Standards and directives

  • ISO 15883 – Washing and disinfecting machines
  • 2014/53/EU – Radio Equipment Directive – RED

Description

The EBI 12-TC230 is an accurate and reliable data logger that provides temperature and conductivity measurements. It is a simple, easy-to-use device that is suitable for use in a wide range of environments and can be used to monitor temperature and conductivity in laboratories, manufacturing plants, warehouses and other environments.

It has a built-in sensor for measuring temperature and conductivity, allowing both parameters to be monitored accurately at the same time. The data is recorded in the device’s internal memory and can be easily transferred to a computer using software available from EBRO. In addition, the EBI 12-TC230 is equipped with a long-lasting battery to ensure a long service life.

The EBI 12 data logger was developed to check the final rinse in the disinfection machine (WD). The measurement takes place during the cleaning programme in the WD. Intermittent restarting of the machine after water withdrawal is not necessary.

When reprocessing medical devices, it is very important to measure the conductivity regularly during the cleaning and disinfection process. The conductivity of the final rinse is an indicator of the quality of the cleaning performance. With this data logger, measurement is possible without interrupting the WD programme.

 

Who is the EBI 12-TC230 suitable for?

It is suitable for a wide range of users who need accurate temperature and conductivity monitoring in a variety of applications. Some examples of users:

  • Laboratories: can be used to monitor temperature and conductivity in laboratories where accurate monitoring of parameters is important.
  • Production plants: useful for monitoring temperature and conductivity in the production of food, beverages, pharmaceuticals and cosmetics.
  • Warehouses: can be used to monitor temperature conditions in warehouses where accurate monitoring of parameters is important.
  • Transport: useful for monitoring temperature conditions during the transport of sensitive products such as medicines and biological samples.
  • Researchers: the perfect choice for researchers who need accurate temperature and conductivity monitoring in a variety of applications.

 

Technical specifications

  • operating temperature: 0 °C … +125 °C
  • measuring range: 0 … +125 °C, 1 µS/cm … 2.000 µS/cm
  • temperature accuracy: ±0.2 °C
  • conductivity accuracy:
    • ±0.5 µS (1 µS/cm .. 20 µS/cm)
    • ±1.0 µS (20.1 µS/cm .. 100 µS/cm)
    • ±3 % of the measurement value (100.1 µS/cm .. 1500 µS/cm)
    • ±5 % ( (for the remaining measuring range)
  • resolution: 0.01 °C, 0.1 µS/cm
  • data storage: 2x 50,160 measurements
  • Sensor: Pt1000, conduction electrode
  • Sample interval: 1 s … 24 h
  • how to measure:
    • infinite measurement
    • measurement at start time
    • instant start to the end of memory
    • start/stop measurement
  • storage temperature: 0 °C … +125 °C
  • battery: lithium 3.6 V, replaceable
  • Dimensions: 48 mm x 130 mm (depending on model)
  • weight: approx. 150 g
  • housing: 316L stainless steel/PEEK
  • Protection class: IP68
  • factory calibration certificate:
    • 0 °C, +60 °C and +121 °C
    • 5.0 µS/cm, 100.0 µS/cm and 1413.0 µS/cm

 

Standards and directives

  • ISO 15883 – Washing and disinfecting machines
  • 2014/53/EU – Radio Equipment Directive – RED

 