EA2 Easidew dew point transmitter in compressed air Michell

Iskanje

Šifra: EA2 Easidew


Po naročilu. Pošljite povpraševanje.

Delivery time: approximately 3-4 weeks

  • temperature range: -110 … +20 °Cdp (°C dew point)
  • Accuracy: ±2 °Cdp
  • response time: 5 min to T95 (dry to wet)
  • operating temperature: -40 … +60 °C
  • temperature compensation range: -20 … +50 °C
  • repeatability: 0.5 °Cdp
  • traceable 13-point calibration certificate
  • Output signal: 4-20 mA (2-wire connection)
MICHELL


Področja:

Description

EA2 Easidew compressed air dew point transmitter Michell is a humidity transmitter that measures the dew point and moisture content of areas/applications with compressed air. It comes with a range of process connectors and electrical connectors. All Easidew dew point converters are available with a service exchange programme that reduces maintenance costs.

Michell’s advanced moisture meters, based on ceramic moisture sensor technology, are calibrated to international standards and are supplied with a traceable calibration certificate.

The mechanical design enables the Easidew Michell compressed air converter to meet the safety and health requirements of the user. It offers ultra-high process pressure barriers, together with an exceptional level of traceability and quality.

 

EA2 Easidew features

  • Output: dew point or output moisture content in ppmV
  • 2-wire loop power connector
  • Power supply 12 … 28 V DC
  • low running costs and easy maintenance with the sensor exchange programme
  • 316 stainless steel sensor and sample blocks
  • Process connections: 5/8″ UNF, 3/4″ UNF, G 1/2″
  • electrical connectors: miniDIN 43650, M12
  • pure oxygen applications
  • implementations:
    EASIDEW I.S.
    (IECEx, CSA, FM and ATEX certificates)
  • Certificate o calibration

 

Areas of use, applications

  • Compressed air in dryers of Class 1 to 6
  • respiratory air quality
  • hydrogen coolant
  • household gases
  • glove storage boxes
  • welding gases
  • hazardous/risk environments/areas
  • oxygen purity

 

Articles

You are invited to take a look at the articles ”Continuous monitoring in compressed air systems”, ”Additive metal production”, ‘ ‘Breathing air quality for divers”, ‘ ‘Moisture in medical gases”.

Povpraševanje