XTP601 EX oxygen analyser Michell

Iskanje

Šifra: XTP601


Po naročilu. Pošljite povpraševanje.

Delivery time: approximately 3-4 weeks

  • measuring range: 0 … 100 % (depending on the gas)
  • Response time: < 50 to < 20 (depending on gas)
  • accuracy: ±0.2 %
  • repeatability: ±0,2 %
  • Response time: < 15 s
  • recommended calibrations: monthly, before use, annually (depending on the gas)
  • measuring technology: thermally conductive sensor
  • ATEX, IECEx, CCSAUS, TC, TR & Ex certificates
MICHELL


Področja:

,

Description

The XTP601 proprietary Michell oxygen analyser for use in safe or hazardous (Ex) areas is a robust, linear and stable device used for the measurement of gases such as biogas, methane (CH4), hydrogen (H), nitrogen (N) or carbon dioxide (CO2). The XTP601 meets the requirements of the IEC61508 SIL2 standard.

It is housed in a rugged IP66 enclosure, making it suitable for a wide range of applications. The oxygen analyser can be supplied with flame retardants, either explosion-proof and certified for use in hazardous areas (Ex zones), or for use in safe areas with flammable gases.

The XTP series offers a basic model XTP601, the specifications of which can be found in the attached documentation. This series also presents
XTP501
for oxygen analysis in safe areas with the same technology, based on a thermo-paramagnetic sensor.

 

Advantages and features of XTP601

  • measurement technology: thermo-paramagnetic oxygen sensor
  • gas: process and non-condensed sample with particles < 5µm
  • internal error (accuracy): better than 1% of the scale
  • zero spread stability: better than 0.25% of spread per month
  • input: external sensor
  • outputs: RS485 and two 4-20 mA outputs (according to NAMUR NE 44 standard)
  • Power supply: 24 V DC
  • meets the requirements of IEC61508 SIL 2
  • touch screen allows calibration and adjustment without the need for a hot work permit
  • low cost due to minimal maintenance, no consumable parts
  • high sensor stability allows minimal calibration costs
  • reliable long-term performance
  • inexhaustible technology
  • easy set-up

 

Areas of use, applications

  • Hydrogen coolant in electric turbines
  • hydrogen production by electrolysis
  • product quality in air separation plants
  • synthetic gas production
  • helium processing
  • fuel cell research
  • regeneration of catalytic converters
  • Air and Argon applications for product quality in double glazing
  • Inert gases for the pharmaceutical and chemical industries
  • biofuels, landfills, recovery plants
  • Furnace control in the steel industry
  • N2 generators
  • O2 generators

 

Software

All XTP601 functions can be controlled using the software on the link.

Povpraševanje